Surface Plasmon Resonance
- Introduced by
- Erich Kretschmann
- Year
- 1971
- Type
- Optical technique
- Family
- Optical Biosensing
What does it do?
Surface Plasmon Resonance (SPR) is a real-time, label-free technique for detecting and monitoring biomolecular interactions at a sensor surface by measuring changes in the refractive index caused by ligand binding.
Developed by Kretschmann in 1971 and applied to biosensing by Liedberg, Nylander, and Lundström in 1983, SPR is now a gold standard for measuring binding kinetics (association and dissociation rates) and equilibrium binding constants in protein interactions, antibody-antigen recognition, and drug discovery.
The intuition behind the method
Light reflected from a metal surface exhibits resonance at a specific angle when the energy couples to collective electron oscillations (plasmons) at the metal-solution interface. When molecules bind to receptors on the metal surface, the refractive index near the surface increases, shifting the resonance angle. By monitoring this angle shift over time, one directly measures binding events in real time without labels. The height and slope of the refractive index change are proportional to ligand mass and binding rate.
History
Kretschmann's 1971 theoretical work established the physics of surface plasmon resonance. Practical biosensing emerged from Liedberg's 1983 demonstration of detecting small molecules and proteins. By the 1990s, Biacore (developed by Johnsson and colleagues at Pharmacia) became the commercial standard, revolutionizing drug discovery by enabling rapid kinetic screening. Modern SPR instruments (Biacore, Reichert Technologies, IBIS) are fully automated and integrated with sample handling and data analysis, making SPR routine in pharmaceutical and biotechnology laboratories.
When to use it
Apply SPR to measure binding kinetics and affinity for any biomolecular interaction: protein-ligand, protein-protein, antibody-antigen, or cell-cell interactions. SPR requires immobilization of one binding partner (the analyte) on the sensor surface; this may affect kinetics compared to solution measurements. SPR is ideal for screening hundreds of ligands, optimizing binding, and validating structural models.
Limitations and considerations
- Requires immobilization of one binding partner; attachment chemistry may impair binding or introduce steric effects
- Nonspecific adsorption to the sensor surface can confound results
- Low molecular weight ligands (< 500 Da) produce small refractive index changes, reducing sensitivity
- Baseline drift and sensor aging affect long-term reproducibility
Source conditions do not replace independent scientific review or validation by an analysis engine.
Strengths
- Label-free: no fluorescent, radioactive, or enzymatic labels needed; measures refractive index change directly
- Real-time kinetics: captures both association (ka) and dissociation (kd) rates in a single experiment
- High-throughput capable: multiple experiments in parallel using sensor arrays or multiplex formats
- No sample consumption: analyte solution passes over the chip and can be collected if needed
Applications
- Drug lead optimization: ranking candidate molecules by binding kinetics and affinity to enable rapid structure-activity relationship (SAR) studies
- Therapeutic antibody characterization: measuring binding affinity and kinetics to native and mutant antigens to guide antibody engineering
- Vaccine development: monitoring antibody-antigen binding post-immunization to validate vaccine efficacy
- Diagnostic assay validation: confirming that antibodies used in clinical assays maintain sufficient specificity and affinity
How it works
How it works
A ligand solution is passed over a sensor chip coated with immobilized receptor molecules (e.g., proteins or antibodies).
Polarized light is directed at the sensor surface at various angles; a surface plasmon resonance occurs at a specific angle.
Binding of ligand to receptor increases refractive index; the resonance angle shifts, detected by a photodiode array.
Sensor response (measured in resonance units, RU) is plotted versus time as ligand is injected, binds, and then dissociates.
Real-time kinetics extract the association rate (ka) and dissociation rate (kd); equilibrium constant Kd = kd/ka; steady-state analysis also yields Kd directly.
Common pitfalls
- Comparing SPR kinetics to solution kinetics without accounting for surface effects, which can slow dissociation rates
- Neglecting mass transport limitations: at high analyte concentrations, diffusion to the surface limits observed binding rate
- Over-interpreting binding affinity from weakly binding ligands: very weak interactions may not reach equilibrium within the experiment
Frequently asked questions
Why can SPR measure binding kinetics directly?
SPR detects binding and unbinding events in real time by monitoring refractive index changes at the sensor surface. The association rate (ka) is evident from the rising phase when ligand is injected; the dissociation rate (kd) is evident from the falling phase when ligand is removed. The true equilibrium constant Kd = kd/ka combines both.
What is the difference between kinetic and steady-state binding constant determination?
Kinetic Kd = kd/ka is extracted from the ratio of dissociation and association rates. Steady-state Kd is determined by fitting binding at equilibrium (plateau phase) versus ligand concentration. Both methods yield Kd, but kinetic analysis provides additional information about mechanism.
How does surface immobilization affect SPR results?
Immobilization can slow dissociation rates (kd) because the receptor is restricted in motion; solution kinetics often show faster dissociation. Association rates (ka) may also be affected by surface crowding and steric effects. Careful surface chemistry and low receptor density minimize these artifacts.